• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL OPTICS TECNIS ITEC PRELOADED 1-PIECE IOL; MONOFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL OPTICS TECNIS ITEC PRELOADED 1-PIECE IOL; MONOFOCAL IOLS Back to Search Results
Model Number PCB00
Device Problem Delivery System Failure (2905)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/06/2016
Event Type  malfunction  
Manufacturer Narrative
The preloaded insertion system was received packaged in the original box.The plunger component was observed in the advanced position.The cartridge was observed correctly engaged into lower body of the pcb00 device.A portion of the lens was received in a plastic bag.Visual inspection at 10x microscope magnification was performed.The remaining portion of the lens was observed stuck at the cartridge tip.The manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specifications.A search revealed that no additional complaints for this order number have been received.The directions for use (dfu) were reviewed.The dfu adequately provide instructions and precautions along with warnings for the proper use and handling of the device.The reported issue of partially delivered was not verified however the issue as staging issues was verified.As a result of the investigation, there is no indication of a product quality deficiency.All pertinent information available to abbott medical optics has been submitted.
 
Event Description
It was reported the intraocular lens (iol) would not come out of the loader/inserter correctly.The iol had patient contact.No vitrectomy was performed.No patient complications were reported.No additional information was provided to abbott medical optics.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS ITEC PRELOADED 1-PIECE IOL
Type of Device
MONOFOCAL IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA
Manufacturer (Section G)
ABBOTT MEDICAL OPTICS INC.
road 402 north, km 4.2
anasco industrial park, pob 14
anasco PR 00610
Manufacturer Contact
pamela mcclain
1700 east st. andrew place
santa ana, CA 92705
7142478243
MDR Report Key6074775
MDR Text Key59063802
Report Number2648035-2016-01738
Device Sequence Number1
Product Code HQL
UDI-Device Identifier05050474558205
UDI-Public(01)05050474558205(17)190727
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P980040
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other
Type of Report Initial
Report Date 11/02/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/02/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date07/27/2019
Device Model NumberPCB00
Device Catalogue NumberPCB0000170
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/19/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received10/06/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/27/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-