Siemens filed the initial mdr 1219913-2016-00203 on 11/03/2016 for (b)(6) advia centaur cp hav total (ahavt) results on two patient samples.On 01/25/2017 - additional information: the cause for the (b)(6) advia centaur cp hav total (ahavt) results is unknown.The two patient samples were provided for additional testing.These samples were tested with three lots of advia centaur cp ahavt reagent (lots 073181, 073182, and 073183), and the results were (b)(6).The reagent readypack in question was not provided for additional testing.Siemens was unable to confirm the customer's observation of (b)(6) results.No conclusion can be drawn the instrument is performing within specification.No further investigation is required.
|