Catalog Number 0250080618
Device Problems Crack (1135); Peeled/Delaminated (1454)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/11/2016
Event Type  malfunction  
Manufacturer Narrative
The product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.Alleged failure: cracked/peeling insulation at tip of shaft.Confirmed failure: cracked/peeling insulation at tip of shaft.Probable root cause: poor autoclave reliability, incorrect sterilization/reprocessing procedure, handling procedures, contact forces, product used beyond defined useful life.The device manufacture date is not known.(b)(4).
 
Event Description
It was reported cracked and peeling insulation at tip of shaft.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG, 10MM PEEK MONOPOLAR HANDLE
Type of Device
ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
kimberly lynch
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key6079158
MDR Text Key59340663
Report Number0002936485-2016-01032
Device Sequence Number1
Product Code OCZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K973259
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 10/11/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/04/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number0250080618
Device Lot Number¿1146661D¿
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/11/2016
Date Manufacturer Received10/11/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-