Model Number PVS25
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem No Information (3190)
Event Date 10/05/2016
Event Type  Injury  
Event Description
The manufacturer was notified on 05-oct-2016 of the following: the patient is a (b)(6) male who had been treated for endocarditis, having avr, mvp, cabg x3 using lima and right greater saphenous vein.Intra-op annulus was measured at 22mm and no stj measurement was taken to figure out a ratio.Three distal graft were performed before the valves.The surgeon open the left atrium and placed 28mm carbiomedics annuloflo ridge ring on the mitral then closed the left atrium.Next he opened the aorta to debride the leaflets and inspect the annulus.There were no abscesses or damage to the annulus from the endocarditis.He started sizing and felt that the large size valve was correct.Valve was prepared according to ifu with no problems while the surgeon placed the three guide sutures 2mm below the nadir.Valve was lowered and checked before deployed.Valve looked good and was balloon at 4atms for 30 seconds with warm saline.Valve was inspected again after ballooning and no issues were seen.The aortotomy was closed and two proximals were done.Crossclamp was removed and heart de-aired.First look at echo the valve had migrated into the valsalva.(i was able capture this image).The crossclamp was then reapplied and heart stopped to remove the valve.The perceval 25/l was explanted and a perceval 23/m was used.Total bypass time 212 minutes.Total cross clamp time: 180 minutes.First cross clamp time: 105 minutes.
 
Manufacturer Narrative
Design history record review completed mar 31, 2017.The complete manufacturing and material records for the perceval heart valve, model icv1210, s/n (b)(4), were pulled and reviewed by quality control at livanova (b)(4) corp.The results confirmed that this valve satisfied all material, visual, and performance standards required for a (model icv1210) perceval heart valve at the time of manufacture and release.
 
Manufacturer Narrative
Conclusion codes, follow-up report #1 did not include conclusion code for user error.The likely root cause of this event with the perceval valve size large is due to mis-sizing the valve.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERCEVAL SUTURELESS AORTIC HEART VALVE
Type of Device
TISSUE HEART VALVE
Manufacturer (Section D)
LIVANOVA CANADA CORP.
5005 north fraser way
burnaby, bc
Manufacturer (Section G)
LIVANOVA CANADA CORP.
5005 north fraser way
burnaby, bc
Manufacturer Contact
francesca crovato
5005 north fraser way
burnaby, bc 
4125650
MDR Report Key6079288
MDR Text Key59186233
Report Number3004478276-2016-00133
Device Sequence Number1
Product Code DYE
UDI-Device Identifier00896208000436
UDI-Public(01)00896208000436(240)ICV1210(17)190319
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 05/26/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/04/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date03/19/2019
Device Model NumberPVS25
Device Catalogue NumberICV1210
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/26/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/19/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age70 YR
Patient Weight97
-
-