MAUDE Adverse Event Report: DEPUY SYNTHES POWER TOOLS 11CM ANGLE ATTACHMENT, BLACK MAX; DRILLS, BURRS, TREPHINES & ACCESSORIES - ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES POWER TOOLS 11CM ANGLE ATTACHMENT, BLACK MAX; DRILLS, BURRS, TREPHINES & ACCESSORIES - ATTACHMENT Back to Search Results
Catalog Number B-QD11
Device Problems Entrapment of Device (1212); Temperature Problem (3022)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/05/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).(b)(6).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If additional information should become available, a supplemental medwatch report will be sent accordingly.
 
Event Description
It was reported from (b)(6) that the attachment device became unusually warm; and that the cutter device could not be removed out of the attachment device.It was not reported if the device was used in surgery, or if there was patient involvement reported.It was not reported if there were any delays in a scheduled surgical procedure or if a spare device was available for use.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
The actual device was returned for evaluation.Reliability engineering evaluated the device and observed that the locking mechanism was not functioning, color marking sleeve was missing (1 ring), tool is blocked in the attachment, sleeve and drive shaft are worn out and screw of sleeve are loose.It was also noted that the device failed pre-test for visual assesment, thimble set screw and cutter insertion.Therefore, the reported condition was confirmed.The assignable root cause was determined to be wear from normal use and servicing.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
11CM ANGLE ATTACHMENT, BLACK MAX
Type of Device
DRILLS, BURRS, TREPHINES & ACCESSORIES - ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
rodney crain
4500 riverside drive
palm beach gardens, FL 33410
5616271080
MDR Report Key6079627
MDR Text Key59574486
Report Number1045834-2016-13140
Device Sequence Number1
Product Code HBE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K002903
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/21/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/04/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberB-QD11
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/08/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received11/24/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/07/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-