• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH 1/8" PEG DRILL; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH 1/8" PEG DRILL; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 6541-4-518
Device Problems Break (1069); Fracture (1260); Difficult to Remove (1528); Sticking (1597)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/10/2016
Event Type  malfunction  
Manufacturer Narrative
It was noted that the device is not available for evaluation.Additional information has been requested and if received, will be provided in the supplemental report upon completion of the investigation.
 
Event Description
While drilling tibial baseplate trail, 1/8 drill with stop broke.No adverse consequences to the patient, surgery performed as normal.Also modular 4 in 1 capture getting stuck on cutting block.Not easy to remove.Information submitted was everything from both surgeon and hospital- no further details.
 
Manufacturer Narrative
Reported event: an event regarding crack/fracture involving a triathlon drill bit was reported.The event was not confirmed.Method & results: -device evaluation and results: a visual, functional and dimensional inspection could not be performed as the device was not returned.-medical records received and evaluation: not performed as patient factors did not contribute to the event.-device history review: device history review indicated the devices accepted into final stock from the reported lot were free from discrepancies -complaint history review: there has been no other event for the lot referenced.Conclusions: the event could not be confirmed nor the root cause could be determined because the devices were not returned for evaluation and insufficient medical information was provided.If the devices and/or additional information are received, this investigation will be reopened and re-evaluated.Product surveillance will continue to monitor for trends.
 
Event Description
While drilling tibial baseplate trail, 1/8 drill with stop broke.No adverse consequences to the patient, surgery performed as normal.Also modular 4 in 1 capture getting stuck on cutting block.Not easy to remove.Information submitted was everything from both surgeon & hospital- no further details.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1/8" PEG DRILL
Type of Device
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
jadwiga degrado
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key6080327
MDR Text Key59504033
Report Number0002249697-2016-03468
Device Sequence Number1
Product Code JWH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K143393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/04/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number6541-4-518
Device Lot NumberSC5K01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/22/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/09/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age63 YR
Patient Weight104
-
-