Model Number 0010-2400
Device Problems Bent (1059); Break (1069); Melted (1385); Temperature Problem (3022)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/16/2016
Event Type  malfunction  
Manufacturer Narrative
Product evaluation: failure analysis for fiber (b)(4): the glass cap exhibits severe devitrification at output window; the metal cap exhibits moderate detritus adhesion on surface, and indications of slight melting on output window; the fiber is broken within the outer flow tubing at 2 inches from control knob, between distal tip and control knob, and exhibits indication of bending, crushed, and slight melting at break location; the fiber was tested with hene laser fixture, aim beam is present at break location; the outer flow tubing open end exhibits minor scratch marks, and mild contamination, likely biologic.Based on device analysis, the potential for forward firing may exist.Probable root cause: based on the device analysis, the probable root cause of the failure is: heat accumulation.Cap wear was accelerated due to anatomical/procedural factors (tissue contact and technique) encountered during the procedure which would limit the performance of the fiber.
 
Event Description
It was reported that during a prostate procedure at 178,492 joules of use and 18 minutes, "fiber broke" was observed.It was further reported that there was a "breakage at insertion site of working element." the fiber tip (cap) was not cleaned during the procedure.The fiber was exchanged and the procedure was completed utilizing a second fiber.Patient's outcome "ok" reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GREENLIGHT MOXY FIBER OPTIC
Type of Device
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE)
3070 orchard drive
san jose CA 95134
Manufacturer (Section G)
AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE)
3070 orchard drive
san jose CA 95134
Manufacturer Contact
jennifer mascioli-tudor
3070 orchard drive
san jose, CA 95134
4084563300
MDR Report Key6081826
MDR Text Key59354118
Report Number2937094-2016-01079
Device Sequence Number1
Product Code GEX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Reporter Occupation Other
Type of Report Initial
Report Date 09/23/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/07/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2018
Device Model Number0010-2400
Device Lot Number622A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/18/2016
Date Manufacturer Received10/27/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-