WILLIAM COOK EUROPE COOK CELECT® PLATINUM NAVALIGN JUGULAR VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Catalog Number IGTCFS-65-1-JUG-CELECT-PT
Device Problem Difficult or Delayed Positioning (1157)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/22/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Catalog # igtcfs-65-1-jug-celect-pt.(b)(4).Summary of investigational findings: investigation of the returned product could not explain the reported "the filter hook would not disengage after several attempts".The root cause is not possible to determine.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended.Cook medical will continue to monitor for similar events.
 
Event Description
Description of event according to initial reporter: the physician attempted to place the ivc filter via jugular access.The filter was unsheathed and attempted to deploy using the appropriate buttons; however the hook would not disengage after several attempts.After unsuccessful attempts, the filter was re sheathed and removed the deployment handle and filter in tact leaving the sheath in place and opened a new filter and successfully placed the filter as intended.Patient outcome: according to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COOK CELECT® PLATINUM NAVALIGN JUGULAR VENA CAVA FILTER SET
Type of Device
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
4632 bjaeverskov
DA 
Manufacturer Contact
marianne hoey
sandet 6
bjaeverskov 4632
DA   4632
56868686
MDR Report Key6082485
MDR Text Key59357158
Report Number3002808486-2016-01331
Device Sequence Number1
Product Code DTK
UDI-Device Identifier10827002343099
UDI-Public(01)10827002343099(17)180923(10)E3369776
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K121629
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 07/26/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/07/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberIGTCFS-65-1-JUG-CELECT-PT
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received07/26/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/23/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age51 YR
-
-