The device, used in treatment, was returned for evaluation.There was a relationship found between the returned device and the reported incident.Visual inspection under magnification shows partial detachment of the electrodes with jagged edges on the remaining electrode leg.There is a significant amount of discoloration on the adhesive around the spacer.There is scratch marks present on the exposed shaft near the tip and the shaft has been bent midway from the tip and handle of the device.The cable has been severed prior to being received for evaluation; therefore, a functional test could not be conducted.There are no manufacturing abnormalities visually observed with the returned device.The complaint has been visually verified and the root cause was more than likely associated with the device potentially coming into contact with a metal object such as a cannula.Other factors that could have led to tip detachment includes: mechanical displacement of tissue through applied force or using the device as a lever to enlarge a surgical site or gain access to tissue.The instruction for use (¿ifu¿) outlines warnings and precautionary measures to adhere to during activation of the device.There were no indications during manufacturing record review that would suggest that the device did not meet product specifications upon release into distribution.
|