Visual evaluation of the returned speedlock hip device shows the implant was not fully deployed and the anchor broke off.The snare line and spring were reeled in with the clinical suture attached.Functional testing could be performed as the speedlock is a single use device.The customer¿s complaint has been visually verified, however, the root cause could not be determined.Potential factors unrelated to the manufacture or design of the device which may have contributed to the reported event includes: levering the device during insertion.It is possible that during or after insertion of the device, if the device is levered, the implant can break off prematurely.The instructions for use (ifu) outlines warnings and precautionary measures when using the device such as ¿use care to properly align the implant and inserter handle with the bone hole during insertion.Do not bend or twist the inserter handle during and after insertion, as damage to the implant or incomplete insertion may result.There were no indications during manufacturing record review that would suggest that the device did not meet product specifications upon release into distribution.
|