Model Number HT70 |
Device Problem
Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The ventilator was returned for evaluation.The service engineer evaluated the device and verified the reported complaint.When the device was powered on, the display froze at the six picture screen.The single board computer (sbc) printed circuit board (pcb) was replaced and the device passed all testing.The reported event was duplicated.
|
|
Event Description
|
It was reported that the ventilator display froze at the six picture screen.There was no patient involvement.
|
|
Manufacturer Narrative
|
The device was repaired and the reported issue was isolated to interface between the device and the failed component.If information is provided in the future, a supplemental report will be issued.
|
|