Catalog Number 111140
Device Problems Break (1069); Detachment Of Device Component (1104); Fracture (1260); Material Integrity Problem (2978)
Patient Problems No Consequences Or Impact To Patient (2199); Foreign Body In Patient (2687)
Event Date 10/10/2016
Event Type  malfunction  
Manufacturer Narrative
Additional information has been requested.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.
 
Event Description
While removing the trials the femoral trial impactor broke off.All pieces were retrieved and no harm was done to the patient.
 
Manufacturer Narrative
An event regarding crack/fracture involving a femoral trial slaphammer was reported.The event was confirmed.Method & results: device evaluation and results: the device was returned in used condition.Examination of the returned device with material analysis engineer indicated that the impactor fractured due to an overload condition.Medical records received and evaluation: not performed as patient factors did not contribute to the event.Device history review: device history review indicated the devices accepted into final stock from the reported lot were free from discrepancies.Complaint history review: there has been 1 other event for the lot referenced.Conclusions: the investigation concluded that the device fractured due to an overload condition.Examination of the returned device with material analysis engineer indicated that the impactor fractured due to an overload condition.
 
Event Description
While removing the trials, the femoral trial impactor broke off.All pieces were retrieved and no harm was done to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMORAL TRIAL SLAPHAMMER
Type of Device
PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer (Section G)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer Contact
cindy chuhinko
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key6085583
MDR Text Key59866378
Report Number3005738134-2016-00002
Device Sequence Number1
Product Code NPJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K090763
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/15/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/08/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number111140
Device Lot Number06030614
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/08/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received01/20/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-