Model Number 407200
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Cardiac Perforation (2513)
Event Date 10/18/2016
Event Type  Injury  
Manufacturer Narrative
(b)(4).The results of the investigation are inconclusive since the device was not returned for analysis.Our investigation was limited to the review of the device history record, which showed that each manufacturing and inspection operation was performed and indicated complete in accordance with sjm specifications and procedures.Based on the information received, the cause of the reported cardiac perforation was procedure related.Per the ifu, cardiac perforation is a known risk during the use of this device.
 
Event Description
During a pulmonary vein isolation ablation procedure, a cardiac perforation occurred.A double transseptal puncture was performed and both introducers were placed to access the left atrium.Following ablation, the patient became hypotensive and an echocardiogram revealed a pericardial effusion for which a pericardiocentesis was performed to stabilize the patient.There were no performance issues with any sjm device during the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRK¿ TRANSSEPTAL NEEDLE, 71 CM LENGTH
Type of Device
TRANSSEPTAL NEEDLE
Manufacturer (Section D)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6087632
MDR Text Key59450711
Report Number3008452825-2016-00152
Device Sequence Number1
Product Code DRC
Combination Product (y/n)N
Reporter Country CodePO
PMA/PMN Number
K072278
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 10/18/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/08/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator No Information
Device Expiration Date07/31/2019
Device Model Number407200
Device Lot Number5603596
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/18/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/11/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Treatment
FLEXABILITY ABLATION CATHETER; INQUIRY AFOCUS DIAGNOSTIC CATHETER; SWARTZ SL INTRODUCER X2
Patient Outcome(s) Other; Required Intervention;
Patient Age60 YR
Patient Weight90
-
-