HILL-ROM BATESVILLE PROGRESSA FRAME; A/C POWERED ADJUSTABLE HOSPITAL BED Back to Search Results
Model Number P7500A000181
Device Problem Device Stops Intermittently (1599)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/11/2016
Event Type  malfunction  
Manufacturer Narrative
The hill-rom technical support representative performed troubleshooting over the phone with the account, asking the account to inspect the patient right side brake activation limit switches for proper operation.It is necessary for the progressa" bed to have an effective maintenance program.We recommend that you do annual preventive maintenance.Examine the brakes to see whether the bed moves when the brakes are set.Replace as necessary.Examine the steering mechanism.Replace or adjust the steering control elements of the steering caster if necessary.Replace the caster if necessary.Troubleshoot in the event of a malfunction.A search of the hill-rom maintenance records did not show hill-rom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.Hill-rom technical support contacted the customer two additional times trying to obtain the resolution for this alleged issue.No response has been received from the account.Based on this information, no further action is required.
 
Event Description
Hill-rom technical support received a report from the account stating the brake not set alarm was working intermittently (cutting in and out).The bed was located in the bed shop at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRESSA FRAME
Type of Device
A/C POWERED ADJUSTABLE HOSPITAL BED
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
lori daulton
1069 state route 46 east
batesville, IN 47006
8129312280
MDR Report Key6089760
MDR Text Key59887076
Report Number1824206-2016-00480
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K122473
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 10/11/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/09/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other Caregivers
Device Model NumberP7500A000181
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/11/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/03/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-