The hill-rom technical support representative performed troubleshooting over the phone with the account, asking the account to inspect the patient right side brake activation limit switches for proper operation.It is necessary for the progressa" bed to have an effective maintenance program.We recommend that you do annual preventive maintenance.Examine the brakes to see whether the bed moves when the brakes are set.Replace as necessary.Examine the steering mechanism.Replace or adjust the steering control elements of the steering caster if necessary.Replace the caster if necessary.Troubleshoot in the event of a malfunction.A search of the hill-rom maintenance records did not show hill-rom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.Hill-rom technical support contacted the customer two additional times trying to obtain the resolution for this alleged issue.No response has been received from the account.Based on this information, no further action is required.
|