Catalog Number LX205
Device Problems Break (1069); Component Falling (1105)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/12/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that during an unknown procedure, the clips fell off and bleeding occurred.The cartridges were broken when the clip applier was introduced.It is unknown how the procedure was completed.No adverse outcome and no other reports were done.
 
Manufacturer Narrative
(b)(4).Batch # 0702a5757056: the analysis results found that the lx205 device was received in good visual condition.In an attempt to replicate the reported incident, the device was tested for functionality.Upon functional testing of the device, the instrument loaded, retained and deployed 6 clips as intended.The instrument was fully functional and conforming to our manufacturing requirements.No conclusion could be reached as to what may have caused the reported incident.The batch history records were reviewed and certed by external manufacturing that the manufacturing criteria was met prior to the release of the equipment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGACLIP CLIP APPLIER - MEDIUM - 14.6 CM
Manufacturer (Section D)
ETHICON ENDO SURGERY, INC (CINCINNATI)
4545 creek rd
cincinnati OH 45242 2803
Manufacturer (Section G)
POST SURGICAL MANUFACTURING
25 plant ave
hauppauge NY 11788 3804
Manufacturer Contact
milton garrett
4545 creek road ml 120a
cincinnati, OH 45242
5133378865
MDR Report Key6095598
MDR Text Key60087537
Report Number1527736-2016-00046
Device Sequence Number1
Product Code GDO
Combination Product (y/n)N
Reporter Country CodeSW
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/12/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/11/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberLX205
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/14/2016
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received11/23/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/28/2007
Is the Device Single Use? No
Type of Device Usage N
Patient Sequence Number1
-
-