Model Number 06002-CP-110
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Code Available (3191)
Event Date 10/17/2016
Event Type  malfunction  
Manufacturer Narrative
Haemonetics has requested that the rotors be returned for evaluation, however, the customer has not responded to the requests.This issue of anticoagulant (ac) depletion has been investigated under a corrective action.The results of that investigation determined the likely cause was the harsh cleaning solution used to clean the pump rollers at the customer site can cause damage to the pump rollers which may lead to a device malfunction.New rotors were sent to the customer for the on-site technician to replace.Device not returned.
 
Event Description
Haemonetics received a complaint on 10/17/2016 for a report of anticoagulant (ac) depletion no donor reaction reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PCS®2 PLASMA COLLECTION SYSTEM
Type of Device
SEPARATOR, AUTOMATED, BLOOD CELL, DIAGNOSTIC,,
Manufacturer (Section D)
HAEMONETICS CORP
400 wood rd
braintree MA 02184
Manufacturer (Section G)
HAEMONETICS CORP
400 wood rd
braintree MA 02184
Manufacturer Contact
julie smith, rn
400 wood rd
braintree, MA 02184
7819170643
MDR Report Key6099389
MDR Text Key60305210
Report Number1219343-2016-00078
Device Sequence Number1
Product Code GKT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
BK920039
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial
Report Date 10/17/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/14/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Phlebotomist
Device Model Number06002-CP-110
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received10/17/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/09/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberB-0349-2016