MAUDE Adverse Event Report: ABBOTT MEDICAL OPTICS EMERALD; SURGICAL ADJUNCTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL OPTICS EMERALD; SURGICAL ADJUNCTS Back to Search Results
Model Number EMERALDC30
Device Problems Difficult or Delayed Positioning (1157); Positioning Failure (1158)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/19/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(6).All pertinent information available to abbott medical optics has been submitted.
 
Event Description
It was reported by the customer that the intraocular lens (iol) got stuck in the injector.The iol was deformed.The product had contact with the patients eye.No patient injury was reported.No additional information was provided to abbott medical optics.
 
Manufacturer Narrative
Device evaluation: the cartridge was not returned for evaluation.The reported issue was not verified.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specifications.Labeling review: the directions for use (dfu) were reviewed.The dfu adequately provide instructions and precautions along with warnings for the proper use and handling of the device.Conclusion: as a result of the investigation, there is no indication of a product quality deficiency.All pertinent information available to abbott medical optics has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMERALD
Type of Device
SURGICAL ADJUNCTS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA
Manufacturer (Section G)
ABBOTT MEDICAL OPTICS INC.
road 402 north, km 4.2
anasco industrial park, pob 14
anasco PR 00610
Manufacturer Contact
pamela mcclain
1700 east st. andrew place
santa ana, CA 92705
7142478243
MDR Report Key6100961
MDR Text Key59851285
Report Number2648035-2016-01796
Device Sequence Number1
Product Code KYB
UDI-Device Identifier05050474530027
UDI-Public(01)05050474530027(17)170509(10)CB31997
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
K961242
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/08/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/14/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date05/09/2017
Device Model NumberEMERALDC30
Device Catalogue NumberEMERALDC30
Device Lot NumberCB31997
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/13/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/09/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
IOL, MODEL AR40E 2807751606
-
-