GE OEC MEDICAL SYSTEMS (SLC) 9900; FLUOROSCOPIC X-RAY Back to Search Results
Model Number 9900
Device Problem Loss of Data (2903)
Patient Problem No Patient Involvement (2645)
Event Date 09/29/2016
Event Type  malfunction  
Manufacturer Narrative
A ge service representative performed an onsite investigation.The cine hard disk drive was evaluated and replaced.The system was tested and found to be working as intended and returned to service.
 
Event Description
The customer reported that the system displayed a cine disk unavailable error and could not save and recall cine patient image data.No patient serious injury or death was reported related to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
9900
Type of Device
FLUOROSCOPIC X-RAY
Manufacturer (Section D)
GE OEC MEDICAL SYSTEMS (SLC)
384 wright brothers drive
salt lake city UT
Manufacturer Contact
384 wright brothers drive
salt lake city, UT 
MDR Report Key6102788
MDR Text Key59895055
Report Number1720753-2016-03106
Device Sequence Number1
Product Code OWB
Combination Product (y/n)N
Reporter Country CodeTH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial
Report Date 11/15/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/15/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9900
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/28/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/01/1970
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-