• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B.BRAUN SURGICAL SA HISTOACRYL BLUE 0.5ML; TISSUE ADHESIVES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B.BRAUN SURGICAL SA HISTOACRYL BLUE 0.5ML; TISSUE ADHESIVES Back to Search Results
Model Number 1050044
Device Problem Partial Blockage (1065)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/16/2016
Event Type  malfunction  
Manufacturer Narrative
Reported device not marketed in the u.S., however, similar devices or devices that share components, raw materials, process methods or any other technological characteristics with a medical device that is registered within the u.S.(b)(4).Manufacturing site evaluation: evaluation on-going.Device not returned.
 
Event Description
Country of complaint: (b)(6).It was reported that the glue can not flow through and the tip is blocked.
 
Manufacturer Narrative
Samples received: 2 unopened pouches.Analysis and results: there are no previous complaints of this batch of which we manufactured and distributed (b)(4) units.The four ampoules received have been analysed and the glue in all the samples received has been extracted without problem, the cannula is not blocked.Reviewed the batch manufacturing record, this product had a normal process and the results during the process fulfill usp/ep and b.Braun surgical requirements.Final conclusion: although the results of the samples received fulfil b.Braun surgical specifications, we take note of this incidence in order to assess if new or additional actions are needed.Actions on product distributed of this reference/batch: based on the conclusion derived from investigation, it is not required to make actions in distributed product.Corrective/preventive actions: according to internal procedures, there is no need to establish corrective or preventive actions.Nevertheless, this complaint is recorded for trending analysis to assess if actions are needed in the future.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HISTOACRYL BLUE 0.5ML
Type of Device
TISSUE ADHESIVES
Manufacturer (Section D)
B.BRAUN SURGICAL SA
121 carretera de terrassa
rubi, barcelona 08191
SP  08191
Manufacturer (Section G)
B.BRAUN SURGICAL SA
121 carretera de terrassa
rubi, barcelona 08191
SP   08191
Manufacturer Contact
nicole broyles
615 lambert pointe drive
hazelwood, MO 63042
3145515988
MDR Report Key6103420
MDR Text Key59980051
Report Number2916714-2016-00933
Device Sequence Number1
Product Code MPN
Combination Product (y/n)N
PMA/PMN Number
NARRATIVE
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 03/16/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/15/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1050044
Device Catalogue Number1050044
Device Lot Number215424N2
Was Device Available for Evaluation? No
Distributor Facility Aware Date10/21/2016
Date Manufacturer Received10/17/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-