• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK SE & CO. KG SOLIA S 60; PACER LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK SE & CO. KG SOLIA S 60; PACER LEAD Back to Search Results
Model Number 377179
Device Problems Low impedance (2285); Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/02/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
Patient was interrogated and found an rv lead failure alert on the device.Low impedance measurements in bipolar and around normal measurements in unipolar were noted.Patient is not dependent.This could possibly be an inner insulation breach or fluid in the header.The rep was to get more information on patient's condition, to turn the device back to bipolar and take measurements.Second round of measurements show impedances in normal range.Device remains implanted and the patient will be monitored.
 
Manufacturer Narrative
The device is currently not available for analysis.No conclusion can be drawn at this time.The investigation will be re-opened should additional data become available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOLIA S 60
Type of Device
PACER LEAD
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin D-123 59
GM  D-12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key6105661
MDR Text Key60030443
Report Number1028232-2016-04533
Device Sequence Number1
Product Code NVN
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P950037
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/02/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/16/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number377179
Device Catalogue NumberSEE MODEL NO.
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received12/01/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/26/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age96 YR
-
-