Device Problems
Premature Discharge of Battery (1057); Moisture Damage (1405)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The pump has been returned to animas.Evaluation has not yet been completed.When evaluation is complete a supplemental report will be filed.No conclusion can be made at this time.
|
|
Event Description
|
On (b)(6) 2016, the reporter contacted animas, alleging a power (battery life w/ moisture) issue.This complaint is being reported because the reported issue was not resolved with troubleshooting.There was no indication that the product caused or contributed to an adverse event.
|
|
Manufacturer Narrative
|
The device has been returned and evaluated by product analysis on 1/05/2017 with the following findings: a review of the pump¿s black box data showed evidence of a short battery life illustrated with a 19 count voltage drop on (b)(6) 2016.During visual inspection of the pump, it was observed that the battery compartment was cracked and there was moisture corrosion observed in the battery canister.It was also observed that the bolus button cover was damaged but the bolus button was still operable.A leak test was performed and failed due a battery compartment leak and a bolus button leak.The ¿ez-prime¿ steps were performed correctly and all electrical current draws were within specification.The initial ¿short battery life¿ complaint was not duplicated during the investigation.The pump¿s cover was removed and there was no further moisture ingress or any intermittent condition found to the printed circuit board (pcb) or to the cgm module.
|
|