DEPUY SYNTHES POWER TOOLS EPD 60000 RPM; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 05.001.010 |
Device Problems
Unstable (1667); Mechanical Jam (2983)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The actual device was returned for evaluation.Reliability engineering evaluated the device and the reported condition was confirmed.The assignable root cause was determined to be due to device wear from normal use and servicing.If additional information should become available, a supplemental medwatch report will be sent accordingly.
|
|
Event Description
|
It was reported from (b)(6) that the electric pen drive device had unstable movement (sometimes would stop and sometimes would not stop).The reporter clarified that the possibility of radio disturbance was low since the user was not using electric knife.During an in-house engineering evaluation, it was observed that the motor was seized, jammed and heavy moving.It was that the device this event did not occur during surgery.It was not reported if there were any delays in a scheduled surgical procedure or if a spare device was available for use.It was reported that this device was a veterinary device.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|