Catalog Number 160177
Device Problems Break (1069); Material Integrity Problem (2978)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 10/20/2016
Event Type  malfunction  
Manufacturer Narrative
Additional information has been requested.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.
 
Event Description
Plastic tibial baseplate impactor tip broke in two pieces during tibial baseplate impaction.
 
Manufacturer Narrative
An event regarding crack/fracture involving an mako impactor was reported.The event was confirmed.Method & results: -device evaluation and results: review of device by material analysis engineer indicated: "fracture consistent with overload as indicated by hackles observed on the fracture surface." -medical records received and evaluation: not performed as medical records were not received and patient factors didn't contribute to the event.-device history review: indicated the devices accepted into final stock from the reported lot were free from discrepancies -complaint history review: there has been no other event for the lot referenced.Conclusions: visual inspection shows the impactor tip broke in two pieces.Examination of the returned device in consultation with material analysis engineer concluded that fracture consistent with overload as indicated by hackles observed on the fracture surface.No further investigation for this event is possible at this time.If additional information will be received, this investigation will be reopened and re-evaluated.
 
Event Description
Plastic tibial baseplate impactor tip broke in two pieces during tibial baseplate impaction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIBIAL INLAY IMPACTOR HEAD
Type of Device
PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer (Section G)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer Contact
joann lavatelli
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key6113496
MDR Text Key60690286
Report Number3005738134-2016-00007
Device Sequence Number1
Product Code NPJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K090763
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/19/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/18/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number160177
Device Lot Number12350915
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/04/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received12/22/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/21/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-