• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES POWER TOOLS COLIBRI; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES POWER TOOLS COLIBRI; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT Back to Search Results
Catalog Number 532.001
Device Problems Device Reprocessing Problem (1091); Incorrect Or Inadequate Test Results (2456); Physical Resistance (2578); Battery Problem (2885); Mechanical Jam (2983); Naturally Worn (2988)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The actual device was returned for evaluation.Reliability engineering evaluated the device and the reported condition was confirmed.The assignable root cause was determined to be due to device wear from normal use and servicing.If additional information should become available, a supplemental medwatch report will be sent accordingly.
 
Event Description
It was reported from (b)(6) that the small battery drive device did not work properly.During pre-repair diagnostic assessment, it was observed that the motor was seized, jammed, heavy moving, motor resistance was too high, motor bearings ran rough, battery lock slides were jammed, triggers rough from dirt, handpiece was worn, bearings, tool coupling lock slides and triggers jammed.It was also reported that the device failed pre-test for off/on/on mode.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COLIBRI
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ  CH4437
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ   CH4437
Manufacturer Contact
mark vornheder
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key6113562
MDR Text Key60854968
Report Number8030965-2016-15638
Device Sequence Number1
Product Code HWE
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Service and Testing Personnel
Type of Report Initial
Report Date 10/13/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/18/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number532.001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/26/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received10/26/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/08/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-