COVIDIEN SCD 700 COMPRESSION SYSTEM-EUR; SCD CONTROLLER Back to Search Results
Model Number 295250
Device Problems Charred (1086); Leak/Splash (1354); Electrical Shorting (2926)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.
 
Event Description
It was reported to covidien on (b)(6) 2016 that an issue occurred with a scd controller.The customer states that the unit is leaking.Service found there was a short circuit and the power supply board was charred.
 
Manufacturer Narrative
03/16/2017.An investigation was performed.This complaint has not been confirmed.The actual sample involved in the reported incident was not returned for evaluations.No additional information, pictures or videos were received.Since no sample was returned for examination, it was not possible to evaluate it as part of a comprehensive failure investigation.All quality assurance testing performed during manufacturing was acceptable.The quality assurance review of the visual, physical and dimensional evaluation results indicated that the product met specification requirements.In addition, all device history records (dhr) are reviewed for accuracy prior to product release.No sample or additional information was received for testing and investigation.If the sample is returned in the future, this complaint will be re-opened for further investigation.No trends or triggers have been found, therefore, a corrective or preventive action is not deemed necessary at this time.It must be noted that in-process controls (such as personnel training, incoming quality acceptance testing for raw material, 100% in process visual inspection and visual acceptance sampling are performed in the plant) are in place to prevent nonconforming product from leaving the manufacturing operations.This complaint will be used for tracking and trending purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCD 700 COMPRESSION SYSTEM-EUR
Type of Device
SCD CONTROLLER
Manufacturer (Section D)
COVIDIEN
building # 10, 789 puxing road
shanghai
CH 
Manufacturer (Section G)
COVIDIEN
building # 10, 789 puxing road
shanghai
CH  
Manufacturer Contact
edward almeida
15 hampshire street
mansfield, MA 02048
5084524151
MDR Report Key6114052
MDR Text Key60755985
Report Number3006451981-2016-00588
Device Sequence Number1
Product Code JOW
Combination Product (y/n)N
Reporter Country CodeIS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/14/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/18/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number295250
Device Catalogue Number295250
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received03/16/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-