STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON PKR INSERT X3 #2 LM/RL -9MM; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Catalog Number 5630-G-209
Device Problems Device Operates Differently Than Expected (2913); Material Deformation (2976); Material Integrity Problem (2978); Positioning Problem (3009)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 10/21/2016
Event Type  malfunction  
Manufacturer Narrative
When completed, the evaluation summary will be submitted in a supplemental report.
 
Event Description
Doctor could not lock insert in place.Opened second poly and it seated.
 
Manufacturer Narrative
An event regarding seating/locking issues involving an triathlon insert was reported.The event was confirmed.Method and results: device evaluation and results: examination of the device by the mar engineer indicated "the damage observed on the distal surface is consistent with contact against the baseplate.Explantation damage is also observed." medical records received and evaluation: not performed as patient factors did not contributed to the event.Device history review: device history review indicated the devices accepted into final stock from the reported lot were free from discrepancies complaint history review: there has been no other event for the lot referenced.Conclusions: the reported event was confirmed as per consultation of the returned device with material analysis engineer who indicated that damage observed on the distal surface is consistent with contact against the baseplate.Explantation damage is also observed.
 
Event Description
Doctor could not lock insert in place.Opened second poly and it seated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON PKR INSERT X3 #2 LM/RL -9MM
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
cindy chuhinko
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key6116027
MDR Text Key60559374
Report Number0002249697-2016-03689
Device Sequence Number1
Product Code HRY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K082567
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/09/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/20/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2018
Device Catalogue Number5630-G-209
Device Lot NumberMMDN3D
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/01/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/13/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/13/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age67 YR
Patient Weight88
-
-