• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES POWER TOOLS RECON SAGITTAL SAW WITH KEY FOR TRS; MOTOR, SURGICAL INSTRUMENT, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES POWER TOOLS RECON SAGITTAL SAW WITH KEY FOR TRS; MOTOR, SURGICAL INSTRUMENT, AC-POWERED Back to Search Results
Catalog Number 05.001.240
Device Problem Failure to Shut Off (2939)
Patient Problem No Patient Involvement (2645)
Event Date 11/04/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that during pre-surgery testing, it was discovered that the sagittal saw device would not shut off.There were no delays to the surgical procedure as an identical spare device was available for use.There was no patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
The actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Reliability engineering evaluated the device.A functional assessment was performed and the device was working properly and it passed the pre-repair diagnostic assessment.The reported condition was not duplicated or confirmed.Therefore, an assignable root cause was not determined.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RECON SAGITTAL SAW WITH KEY FOR TRS
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ  CH4437
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ   CH4437
Manufacturer Contact
michael cote
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key6118001
MDR Text Key60876354
Report Number8030965-2016-15662
Device Sequence Number1
Product Code GEY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 11/09/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/21/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number05.001.240
Device Lot Number2457
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/23/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received12/14/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/28/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-