STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON P/A PS BEADED #7L; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number 5516F701
Device Problems Delivered as Unsterile Product (1421); Packaging Problem (3007); Device Handling Problem (3265)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 10/26/2016
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of the investigation.
 
Event Description
Primary surgery of left knee due to pain.Surgeon questioned sterility of the wrapping of the device during case, so it was not implanted.A second device was implanted.
 
Manufacturer Narrative
An event regarding packaging damage involving a triathlon femoral component was reported.The event was confirmed.Method and results: device evaluation and results: based on the visual inspection of the returned packaging it appears that this component packaging was subjected to excessive/inappropriate handling causing excessive movement within the packaging therby causing the device to impact heavily upon the tyvek lid causing the damage noted.Medical records received and evaluation: not performed as the event is related to a packaging issue and no adverse consequences to the patient were reported.Device history review: dhr review for the reported lot determined that the device was manufactured and packed to specification.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: the investigation concluded that the packaging damage was most likely caused by excessive/incorrect handling prior to opening of the packaging.No further investigation for this event is required at this time.No further investigation is possible at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event Description
Primary surgery of left knee due to pain.Surgeon questioned sterility of the wrapping of the device during case, so it was not implanted.A second device was implanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON P/A PS BEADED #7L
Type of Device
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-LIMERICK
raheen business park
limerick NA
Manufacturer Contact
cindy chuhinko
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key6118289
MDR Text Key60622975
Report Number0002249697-2016-03716
Device Sequence Number1
Product Code MBH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K141056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/09/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/21/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2020
Device Catalogue Number5516F701
Device Lot NumberAHT9P
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/10/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/16/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/09/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age55 YR
-
-