Reported event: an event regarding wear involving a triathlon insert was reported.The event was confirmed.Method & results: -device evaluation and results: a visual inspection of the returned device noted wear impression markings on the distal surface of the insert, consistent with contact against the baseplate.Explantation damage was observed on the anterior surface of the insert.Damage was also observed on the posterior surface of the insert.Delamination, burnishing, scratching, third-body indentations and embedded debris was also observed on the condyles.A material analysis was conducted on the received devices and the report concluded "delamination, burnishing, scratching, pitting and third-body indentations were observed on the condyles.These are common damage modes of uhmwpe.Wear impression markings, explantation damages and embedded debris were also observed on the insert.No material or manufacturing defects were observed on the surfaces examined." -medical records received and evaluation: not performed as no medical records were provided.-device history review: indicated the devices accepted into final stock from the reported lot were free from discrepancies.-complaint history review: there has been no other event for the lot referenced.Conclusions: a material analysis was conducted on the received devices and the report concluded "delamination, burnishing, scratching, pitting and third-body indentations were observed on the condyles.These are common damage modes of uhmwpe.Wear impression markings, explantation damages and embedded debris were also observed on the insert.No material or manufacturing defects were observed on the surfaces examined." the event was confirmed but the root cause could not be determined.No further investigation for this event is possible at this time.If the additional information will be received, this investigation will be reopened and re-evaluated.
|