(b)(4).Batch # n9193y.The analysis results found that the mcs20 device was returned with no damage in the external components.In an attempt to replicate the reported incident, the device was tested for functionality.Upon testing, the device was cycled and it fed and it ejected 11 clips.Finally, the instrument locked out.No conclusion could be reached as to what may have caused the reported incident.The batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.
|