Catalog Number K1023 SMN10445160 |
Device Problem
Low Test Results (2458)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/23/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
Analysis of the instrument data indicate that the cause for the falsely depressed qc and patient results is unknown.The low ca results were obtained within a flex reagent cartridge well set.The low results were obtained from a single well set 12:6.The incident is under investigation by the siemens headquarters support center (hsc).
|
|
Event Description
|
Falsely depressed calcium (ca) results were obtained on qc and patient samples on the dimension vista 500 system.Patient results were reported.The same samples were rerun on the same instrument with an alternate reagent well set and higher results were obtained for the qc and patient samples.Corrected patient results were reported.There is no indication that patient treatment was altered or prescribed on the basis of the falsely depressed ca results.There was no report of adverse health consequences as a result of the falsely depressed ca results.
|
|
Manufacturer Narrative
|
Original mdr 2517506-2016-00412 was filed 22-nov-2016.The complaint was unable to be confirmed.
|
|
Search Alerts/Recalls
|