COOK ENDOSCOPY FUSION OMNI-TOME PRE-LOADED SPHINCTEROTOME; KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
|
Back to Search Results |
|
Catalog Number FS-OMNI-35 |
Device Problem
Migration or Expulsion of Device (1395)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/13/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
Concomitant medical products.Erbe electrosurgical generator, unknown model investigation evaluation: due to the condition of the returned device, we were unable to complete a full evaluation.Our laboratory evaluation of the product said to be involved determined that the wire guide lumen of the sphincterotome has been partially utilized.The outer edge of the wire guide lumen exhibits rippling along the separation line and stops approximately 70 cm from the handle.Since the wire guide lumen was returned partially utilized and torn, functional testing with a wire guide separating the wire guide lumen was unable to be performed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: a corrective action has been initiated to reduce occurrences of this nature.The product said to be involved is included in the scope of the corrective actions.Prior to distribution, all fusion omni-tome pre-loaded sphincterotomes are subjected to a visual inspection and functional test to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a corrective action has been initiated in an effort to reduce occurrences of this nature.This product was manufactured prior to implementation of this corrective action.A review of the complaint history was conducted.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
|
|
Event Description
|
During an endoscopic retrograde cholangiopancreatography (ercp), the physician used a cook fusion omni-tome pre-loaded sphincterotome.During a catheterization of the papilla, the user found that the sphincterotome tore during the peeling [exchange] leading to a loss of the way [lost wire guide access].Another sphincterotome had to be used to complete the procedure.
|
|
Search Alerts/Recalls
|
|
|