• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK ENDOSCOPY FUSION OMNI-TOME PRE-LOADED SPHINCTEROTOME; KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK ENDOSCOPY FUSION OMNI-TOME PRE-LOADED SPHINCTEROTOME; KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) Back to Search Results
Catalog Number FS-OMNI-35
Device Problem Migration or Expulsion of Device (1395)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/13/2016
Event Type  malfunction  
Manufacturer Narrative
Concomitant medical products.Erbe electrosurgical generator, unknown model investigation evaluation: due to the condition of the returned device, we were unable to complete a full evaluation.Our laboratory evaluation of the product said to be involved determined that the wire guide lumen of the sphincterotome has been partially utilized.The outer edge of the wire guide lumen exhibits rippling along the separation line and stops approximately 70 cm from the handle.Since the wire guide lumen was returned partially utilized and torn, functional testing with a wire guide separating the wire guide lumen was unable to be performed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: a corrective action has been initiated to reduce occurrences of this nature.The product said to be involved is included in the scope of the corrective actions.Prior to distribution, all fusion omni-tome pre-loaded sphincterotomes are subjected to a visual inspection and functional test to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a corrective action has been initiated in an effort to reduce occurrences of this nature.This product was manufactured prior to implementation of this corrective action.A review of the complaint history was conducted.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event Description
During an endoscopic retrograde cholangiopancreatography (ercp), the physician used a cook fusion omni-tome pre-loaded sphincterotome.During a catheterization of the papilla, the user found that the sphincterotome tore during the peeling [exchange] leading to a loss of the way [lost wire guide access].Another sphincterotome had to be used to complete the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSION OMNI-TOME PRE-LOADED SPHINCTEROTOME
Type of Device
KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
scottie fariole
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key6122329
MDR Text Key60868092
Report Number1037905-2016-00478
Device Sequence Number1
Product Code KNS
UDI-Device Identifier00827002319110
UDI-Public(01)00827002319110(17)190725(10)W3750526
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K052051
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 11/22/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/22/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberFS-OMNI-35
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date10/13/2016
Device Age3 MO
Event Location Hospital
Date Manufacturer Received10/31/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/25/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
FUJINON ENDOSCOPE, UNKNOWN MODEL
-
-