Catalog Number 482317570 |
Device Problems
Detachment Of Device Component (1104); Mechanical Problem (1384); Difficult to Remove (1528)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/28/2016 |
Event Type
malfunction
|
Event Description
|
It was reported that the left screw from lilac bone was being attempted to be removed.Holding the screw head during locked the outer tube of screw driver, attempted to lose the screw, however removed screw head only.Removed the remaining screw part in lilac bone using adjustment driver and complete to remove screw whole.Additional information received from sr by phone on 4-nov-2016.Implanted xia3(ilia) and es2(lumbar) both for fractures on (b)(6) 2015.And removed xia3 as planned on (b)(6) 2016.
|
|
Manufacturer Narrative
|
Method: device history review; complaint history review; risk assessment; results: manufacturing records were reviewed for the corresponding lot and no relevant issues were identified.It was confirmed that the patient fused.Conclusion: the plausible root cause of the xia screw tulip head disengagement is due to patient's fusion and excessive force required to remove the screw.
|
|
Event Description
|
It was reported that the left screw from lilac bone was being attempted to be removed.Holding the screw head during locked the outer tube of screw driver, attempted to lose the screw, however removed screw head only.Removed the remaining screw part in lilac bone using adjustment driver and complete to remove screw whole.Additional information received from (b)(4) by phone on (b)(6) 2016.Implanted xia3(ilia) and es2(lumbar) both for fractures on (b)(4) 2015.And removed xia3 as planned on (b)(6) 2016.
|
|
Search Alerts/Recalls
|