• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER INFINITI VISION SYSTEM; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER INFINITI VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065750833
Device Problems Loose or Intermittent Connection (1371); Output below Specifications (3004)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/25/2016
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A customer reported loosing vacuum during cataract surgery.The case was completed without patient harm.
 
Manufacturer Narrative
The system was examined and the reported event was not replicated.The system was tested and found to meet product specifications.However, as a preventive measure (pm), the connectors were replaced and retuned for evaluation.The system was manufactured on december 18, 2008.Based on qa assessment, the product met specifications at the time of release.A cpc connector was received for evaluation.The returned cpc connector was replaced as part of the system pm.Therefore, the returned cpc connector will not be inspected.The system was found to meet specifications.Therefore, the root cause of the reported event cannot be established.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITI VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
rita lopez
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8175514846
MDR Report Key6128603
MDR Text Key61201950
Report Number2028159-2016-05396
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K120912
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
Report Date 01/31/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/28/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065750833
Other Device ID Number2.06
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/10/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received01/26/2017
Date Device Manufactured12/18/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-