• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN KANGAROO JOEY,PUMP W/CLMP,REFB; ENTERAL FEEDING PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN KANGAROO JOEY,PUMP W/CLMP,REFB; ENTERAL FEEDING PUMP Back to Search Results
Model Number 383401
Device Problems Crack (1135); Overheating of Device (1437)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Submit date: 11/28/2016.An investigation is currently underway.Upon completion, the results will be forwarded.
 
Event Description
It was reported to covidien on (b)(6) 2016 that an issue occurred with the enteral feeding pump.The customer did not state a specific issue on this back to stock unit.Upon triage on (b)(6) 2016, the service tech found the unit had a cracked c144 on the pcb and the board heats up.
 
Manufacturer Narrative
The unit was triaged and the reported issue was confirmed.The root cause is due to a damaged capacitor.The unit was repaired and returned.All device history records are reviewed for quality inspections and parameter compliance prior to releasing the product for shipment.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KANGAROO JOEY,PUMP W/CLMP,REFB
Type of Device
ENTERAL FEEDING PUMP
Manufacturer (Section D)
COVIDIEN
2824 airwest boulevard
plainfield IN 46168
Manufacturer (Section G)
COVIDIEN
2824 airwest boulevard
plainfield IN 46168
Manufacturer Contact
edward almeida
15 hampshire st
mansfield, MA 02048
5084524151
MDR Report Key6129546
MDR Text Key61233760
Report Number3008361498-2016-00178
Device Sequence Number1
Product Code LZH
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/28/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number383401
Device Catalogue Number383401
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received11/21/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-