Model Number 383401 |
Device Problems
Crack (1135); Overheating of Device (1437)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Submit date: 11/28/2016.An investigation is currently underway.Upon completion, the results will be forwarded.
|
|
Event Description
|
It was reported to covidien on (b)(6) 2016 that an issue occurred with the enteral feeding pump.The customer did not state a specific issue on this back to stock unit.Upon triage on (b)(6) 2016, the service tech found the unit had a cracked c144 on the pcb and the board heats up.
|
|
Manufacturer Narrative
|
The unit was triaged and the reported issue was confirmed.The root cause is due to a damaged capacitor.The unit was repaired and returned.All device history records are reviewed for quality inspections and parameter compliance prior to releasing the product for shipment.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.If information is provided in the future, a supplemental report will be issued.
|
|