In the decontamination/complaints laboratory, all four sides of the oxygenator were sawn open, and the mats were inspected for any signs of damage, marks or any other anomalies.No anomalies were found.The therapy application manager evaluated the issue from a clinical perspective.Unfortunately, the clinical information is insufficient to determine a clear root cause.The results of the performance tests confirm to some extent an impaired gas exchange performance at high flow conditions.Although the applied flow rate at the incident site is unknown, it cannot be ruled out that high flow rates were applied.Thrombus formation has been confirmed which could be a possible root cause for impaired blood/insufficient oxygen delivery.It is unclear whether the operator had set the venous saturation low threshold to trigger an alarm on the cardiohelp.However, according to the issue reported by the customer, this is likely not to be the case, and this does not represent the root cause of the issue, but rather a device safety mitigation which might not have been utilized.Maquet cardiopulmonary has no corresponding data about the setup, flow parameters, anticoagulation protocol, the patient's medical history (especially coagulation disorders), etc.In the absence of further information, the root cause remains unclear.
|