COVIDIEN KANGAROO JOEY,PUMP W/POLE CLMP; ENTERAL FEEDING PUMP Back to Search Results
Model Number 383400
Device Problems Improper Flow or Infusion (2954); Pumping Problem (3016)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/12/2016
Event Type  malfunction  
Manufacturer Narrative
Submit date: 11/29/2016.An investigation is currently underway.Upon completion, the results will be forwarded.
 
Event Description
It was reported to covidien on (b)(6) 2016 that an issue occurred with an enteral feeding pump.The customer reports feeding pump is infusing backwards.Additional information has been requested with no response.If additional pertinent information becomes available, the report will be updated.
 
Manufacturer Narrative
An evaluation of the kangaroo pump was performed for the reported condition of infusing backwards.The unit was triaged and the reported issue was not confirmed, but found gearbox rotates in both directions.The gearbox issue is caused by a worn clutch which allows the user to manually turn the rotor in reverse.Even with some wear on the clutch it will continue to engage the shaft so that reverse rotation during operation is prevented.Enhancements have been implemented to prevent the reported condition from recurring.All device history records are reviewed for quality inspections and parameter compliance prior to releasing the product for shipment.Post market trending is in place to monitor for further occurrences.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KANGAROO JOEY,PUMP W/POLE CLMP
Type of Device
ENTERAL FEEDING PUMP
Manufacturer (Section D)
COVIDIEN
2824 airwest boulevard
plainfield IN 46168
Manufacturer (Section G)
COVIDIEN
2824 airwest boulevard
plainfield IN 46168
Manufacturer Contact
edward almeida
15 hampshire st
mansfield, MA 02048
5084524151
MDR Report Key6132648
MDR Text Key61354368
Report Number3008361498-2016-00182
Device Sequence Number1
Product Code LZH
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/27/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/29/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number383400
Device Catalogue Number383400
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received10/14/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-