Model Number 383400 |
Device Problems
Improper Flow or Infusion (2954); Pumping Problem (3016)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/12/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
Submit date: 11/29/2016.An investigation is currently underway.Upon completion, the results will be forwarded.
|
|
Event Description
|
It was reported to covidien on (b)(6) 2016 that an issue occurred with an enteral feeding pump.The customer reports feeding pump is infusing backwards.Additional information has been requested with no response.If additional pertinent information becomes available, the report will be updated.
|
|
Manufacturer Narrative
|
An evaluation of the kangaroo pump was performed for the reported condition of infusing backwards.The unit was triaged and the reported issue was not confirmed, but found gearbox rotates in both directions.The gearbox issue is caused by a worn clutch which allows the user to manually turn the rotor in reverse.Even with some wear on the clutch it will continue to engage the shaft so that reverse rotation during operation is prevented.Enhancements have been implemented to prevent the reported condition from recurring.All device history records are reviewed for quality inspections and parameter compliance prior to releasing the product for shipment.Post market trending is in place to monitor for further occurrences.If information is provided in the future, a supplemental report will be issued.
|
|