• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - ENGLEWOOD EM 2400, DISPLAY MODULE; SYSTEM/DEVICE, PHARMACY COMPOUNDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - ENGLEWOOD EM 2400, DISPLAY MODULE; SYSTEM/DEVICE, PHARMACY COMPOUNDING Back to Search Results
Catalog Number 2400D
Device Problem Filling Problem (1233)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/04/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that there was a set up issue during use of an automated compounding device (acd).The reporter stated that a non-vented inlet had been used to compound a sodium chloride vial, rather than a vented inlet.The compounded bags were within the acceptable weight range and were delivered to the patient.The customer later tested a bag from the compounding batch, and reported that the bag had been compounded with ¿almost 0¿ sodium chloride due to the issue.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
The device was not returned and the lot number is unknown; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EM 2400, DISPLAY MODULE
Type of Device
SYSTEM/DEVICE, PHARMACY COMPOUNDING
Manufacturer (Section D)
BAXTER HEALTHCARE - ENGLEWOOD
englewood CO
Manufacturer (Section G)
BAXTER HEALTHCARE - ENGLEWOOD
14445 grasslands dr
englewood CO 80112
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key6133460
MDR Text Key61085104
Report Number1416980-2016-17743
Device Sequence Number1
Product Code NEP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/29/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2400D
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received02/10/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-