BAXTER HEALTHCARE - ENGLEWOOD EM 2400, DISPLAY MODULE; SYSTEM/DEVICE, PHARMACY COMPOUNDING
|
Back to Search Results |
|
Catalog Number 2400D |
Device Problem
Filling Problem (1233)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/04/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that there was a set up issue during use of an automated compounding device (acd).The reporter stated that a non-vented inlet had been used to compound a sodium chloride vial, rather than a vented inlet.The compounded bags were within the acceptable weight range and were delivered to the patient.The customer later tested a bag from the compounding batch, and reported that the bag had been compounded with ¿almost 0¿ sodium chloride due to the issue.There was no patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
The device was not returned and the lot number is unknown; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
|
|