(b)(4).The external visual inspection revealed that the electrode was severed prior to receipt as well as slices on the fantail region and damage to the epoxy header.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The electrode condition prevented some of the electrical tests from being performed.The device passed the electrical tests performed.The device failed the residual gas analysis test.The device passed the electrical tests performed.However, this device had a moisture that exceeded the residual gas analysis test limit.Based on the residual gas analysis data, it is believed that this device was not hermetic.This older device configuration is no longer manufactured.This is the final report.
|