• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ADVANCED BIONICS, LLC CLARION¿ IMPLANT; COCHLEAR IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADVANCED BIONICS, LLC CLARION¿ IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number AB-5100H
Device Problem Operating System Version or Upgrade Problem (2997)
Patient Problem No Code Available (3191)
Event Date 11/07/2016
Event Type  Injury  
Event Description
The recipient reportedly elected to undergo revision surgery for a technology upgrade.The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.
 
Manufacturer Narrative
(b)(4).The external visual inspection revealed that the electrode was severed prior to receipt as well as slices on the fantail region and damage to the epoxy header.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The electrode condition prevented some of the electrical tests from being performed.The device passed the electrical tests performed.The device failed the residual gas analysis test.The device passed the electrical tests performed.However, this device had a moisture that exceeded the residual gas analysis test limit.Based on the residual gas analysis data, it is believed that this device was not hermetic.This older device configuration is no longer manufactured.This is the final report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLARION¿ IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
Manufacturer Contact
rachael perez`
28515 westinghouse place
valencia, CA 91355
6613627734
MDR Report Key6133890
MDR Text Key61084513
Report Number3006556115-2016-00508
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/09/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/29/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date01/05/2002
Device Model NumberAB-5100H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/09/2016
Is the Reporter a Health Professional? Yes
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/12/2000
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age88 YR
-
-