DEPUY SYNTHES POWER TOOLS BATTERY OSCILLATOR II FOR BPL II; MOTOR, SURGICAL INSTRUMENT, AC-POWERED
|
Back to Search Results |
|
Catalog Number 530.710_LL |
Device Problems
Sticking (1597); Device Stops Intermittently (1599); Naturally Worn (2988)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Additional narrative: (b)(4).The actual device was returned for evaluation.Reliability engineering evaluated the device and observed that the device functioned properly.Therefore, the reported condition was not confirmed.An assignable root cause was not determined.However, it was noted that the triggers were sticking; and that the components were worn.It was determined that this was due to wear from normal use over time.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
It was reported that the battery oscillator device was working intermittently.During an in-house engineering evaluation, it was observed that the triggers were sticking and worn.It was not reported if the device was used in surgery, or if there was patient involvement reported.It was not reported if there were any delays in a scheduled surgical procedure or if a spare device was available for use.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|