Model Number N/A |
Device Problem
Device Operates Differently Than Expected (2913)
|
Patient Problem
No Information (3190)
|
Event Date 11/04/2016 |
Event Type
Injury
|
Manufacturer Narrative
|
Current information is insufficient to permit a conclusion as to the cause of the event.Review of device history records show the lot released with no recorded anomaly or deviation.
|
|
Event Description
|
During a knee arthroplasty the tibia resection block could not be placed in the tibia resecotre depth gauge resulting in an inaccurate cut, which was corrected by the surgeon.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was not confirmed.Review of photo provided did not identify any damage that would prevent the resection block from mating with the depth gauge.Dhr was reviewed and no discrepancies were found.Review of the complaint history determined that no further action is required as no were trends identified.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|