Evaluation of the returned device did not find any evidence of the tip being split or torn on the sides.The reported issue (split tip) was not confirmed.However, the tip was bent was bent and the lens was found stuck in the delivery device tip with both haptics bent.The plunger rod was found to be advanced passed the lens up into the tip.Review of device history record (dhr) indicated no anomalies.Based on available information, a root cause for the reported event could not be conclusively determined.However, operational context due to operator technique such as improper loading of lens may have caused or contributed to the reported issue.
|