• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON LENSX, INC. LENSX LASER SYSTEM; OPHTHALMIC FEMTOSECOND LASER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON LENSX, INC. LENSX LASER SYSTEM; OPHTHALMIC FEMTOSECOND LASER Back to Search Results
Model Number 550
Device Problem Material Fragmentation (1261)
Patient Problem Corneal Clouding/Hazing (1878)
Event Date 11/10/2016
Event Type  Injury  
Manufacturer Narrative
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A surgeon noted hazy corneal appearance after laser assisted portion and prior to cataract removal of the right eye.Surgeon reported capsulotomy, lens fragmentation, primary and secondary incisions were completed on the laser with no complications.Surgeon also indicated there was only one docking and suction attempt made, which was successful.
 
Manufacturer Narrative
There was no technical services requested or performed related to this reported event.With no additional, related information provided, the customer reported event was not able to be confirmed.The device history records (dhr) for the device was reviewed.No abnormalities that could have contributed to this event were found.The associated device was released based on company acceptance criteria.Based on the information obtained, the root cause of the reported event cannot be determined conclusively.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LENSX LASER SYSTEM
Type of Device
OPHTHALMIC FEMTOSECOND LASER
Manufacturer (Section D)
ALCON LENSX, INC.
33 journey
suite #175
aliso viejo CA 92658
Manufacturer (Section G)
ALCON LENSX, INC.
33 journey
suite #175
aliso viejo CA 92658
Manufacturer Contact
rita lopez
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8175514846
MDR Report Key6139698
MDR Text Key61289638
Report Number3008772169-2016-00804
Device Sequence Number1
Product Code OOE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K101626
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 03/02/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/01/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number550
Device Catalogue Number8065998162
Device Lot NumberASKU
Other Device ID Number00380659981623
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/01/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/29/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age73 YR
-
-