Catalog Number 10453571 |
Device Problem
Use of Device Problem (1670)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/29/2016 |
Event Type
No Answer Provided
|
Manufacturer Narrative
|
Siemens issued an urgent field safety notice (# poc 16-019.A.Ous) in june 2016 informing customers of a software defect in the stratus cs acute care diagnostics system regarding testpak calibration and quality control (qc) status when a testpak is due to expire within 48 hours.As stated in the "actions to be taken by the customer" section of the urgent field safety notice: do not use testpaks that are due to expire within 48 hours.The customer acknowledged that they received the urgent field safety notice prior to this event.The customer reported that the testpak lot has been discarded.
|
|
Event Description
|
The customer reported that they calibrated a ctni testpack that was about to expire and got an "*****above assay range" error message when a sample was run.There was no report of injury due to this event.
|
|