• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER, INC. ZIMMER TOTAL HIP MODULAR FLEX DRILL BIT; HIP INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER, INC. ZIMMER TOTAL HIP MODULAR FLEX DRILL BIT; HIP INSTRUMENT Back to Search Results
Model Number N/A
Device Problem Fracture (1260)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Type  malfunction  
Event Description
During a procedure a drill brit fractured while drilling through the drill guide.Fractured pieced were retrieved from the patient and another drill bit of the same size was available to complete the procedure without a significant delay.
 
Manufacturer Narrative
Complaint sample was evaluated and the reported event was confirmed.Device history record was reviewed and no discrepancies were found.Dimensional analysis of the broken drill bit found no deviations from the print specifications.Review of the complaint history determined that no further action is required as no trends were identified.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIMMER TOTAL HIP MODULAR FLEX DRILL BIT
Type of Device
HIP INSTRUMENT
Manufacturer (Section D)
ZIMMER, INC.
1800 west center street
warsaw IN 46580
Manufacturer (Section G)
ZIMMER, INC.
1800 west center street
warsaw IN 46580
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key6142935
MDR Text Key61400133
Report Number0001822565-2016-04525
Device Sequence Number1
Product Code HTW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberN/A
Device Catalogue Number00879000703
Device Lot Number63146492
Other Device ID NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/18/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/04/2016
Initial Date FDA Received12/02/2016
Supplement Dates Manufacturer Received11/09/2017
Supplement Dates FDA Received11/10/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/12/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Other;
-
-