It was reported by a company representative that during a medical procedure, the threaded screw end portion of a bone repositioning device broke off in a patient¿s bone.The broken off piece was removed from the patient.The was no medical intervention.There were no adverse consequences, nor were there any additional symptoms of infections or adverse effects reported.
|
The reported event that the tip of the repositioning instrument broke off could be confirmed according to the provided images via echs.Because the affected instrument was not returned for investigation, a metallurgical examination ¿ indispensable in such cases ¿ cannot be performed making it impossible to determine the exact root cause.However, related to the visual investigation of the provided images the most likely root cause of the breakage could be attributed to an inappropriate user related handling issue.According to that, the repositioning pin was not fully inserted up to the shoulder ring.Related to the determined observations via the provided images the ifu clearly indicates that ¿the repositioning pin must be carefully inserted up to the shoulder ring before repositioning the loose fragments¿ and ¿the instrument is not to be used as a lever.In case of misuse, the instrument could break¿.According to previous conducted investigations further possible root causes include: too high torsional forces during insertion; application of too high bending forces upon to the thread of the pin due to not completely cut off zygoma or other loose bone fragments during repositioning; the instrument was used as a lever; insufficient cleaning caused pitting corrosion ¿ corrosion favours breakages.Related to the submitted production code ag05 the returned device was manufactured after the measurements of capa # 358 were implemented.However, related to the provided images and according to the results within previous conducted investigations the event potentially did not involve a product problem indicating a non-conformity or unanticipated hazard, therefore, no further action is deemed to be required at this time.Product surveillance will continue to monitor complaints of this type for further adverse trends.
|