It was reported by a company representative that the physician's tablet usb port is broken.The cable connected to the tablet felt really loose so the representative tried removing the cable and putting the cable back in.When the cable was inserted back into the tablet, the cable wouldn¿t stay in the tablet.Per the company representative, the usb port was loose within the tablet and fell off.The physician had backup equipment and so they were still able to interrogate the patient and no patient was affected.A replacement tablet was provided to the physician.The suspect tablet was received on 11/22/2016.An analysis was performed on the returned tablet and the reported allegation was verified.During the analysis it was identified that the usb port was damaged.As a result, the tablet was unable to establish communication.No further anomalies were identified.
|