Model Number C12058
Device Problems Fracture (1260); Material Fragmentation (1261)
Patient Problem Vascular Dissection (3160)
Event Date 10/30/2016
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
When the pressurewire aeris was being used with a balloon for pci, the portion of the device outside of the patient became kinked and fractured.When the wire was being manipulated back to the mid rca, it jumped forward and dissected the vessel.Additional stenting was needed to resolve the dissection.
 
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record for the above-referenced product was unable to be reviewed since the batch number is unknown.Based on the information received, the cause of the reported incident could not be conclusively determined.The pressurewire instructions for use (ifu) states that excessive manipulation of the pressurewire when the sensor element or pressurewire tip is located in sharp bend may cause damage or tip fracture.The pressurewire instructions for use (ifu) states that torqueing the pressurewire against resistance or repeated attempts to cross a total vessel occlusion may cause damage and/or fracture, which may lead to a portion of pressurewire separating from the tip.The pressurewire instructions for use (ifu) states that vessel dissection is a potential complication which may be encountered during all catheterization procedures.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRESSUREWIRE¿ AERIS¿
Type of Device
TRANSDUCER, PRESSURE,CATHETER TIP
Manufacturer (Section D)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key6153088
MDR Text Key61696133
Report Number3008452825-2016-00169
Device Sequence Number1
Product Code DXO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131452
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date 11/17/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/07/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberC12058
Device Catalogue NumberC12058
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/17/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-