Model Number 20128E |
Device Problems
Break (1069); Failure to Disconnect (2541)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 11/17/2016 |
Event Type
Injury
|
Manufacturer Narrative
|
Although requested, the affected product has not been received.A follow up report will be submitted with investigation results should the devices be received for evaluation.
|
|
Event Description
|
The customer reported that the tubing broke off inside the central line catheter and the patient had to have their central line replaced.There is no report of lasting patient harm.
|
|
Manufacturer Narrative
|
The affected product has been received and the investigation is pending.A follow up report will be submitted once the failure investigation has been completed.
|
|
Manufacturer Narrative
|
The customer¿s report that the set broke off inside the central line catheter was confirmed.Visual inspection showed the male luer tip was broken off, with stress marks visible at the break site.The broken off male luer tip was found lodged into the female port of the non-bd extension set.Functional testing was not performed due to the broken male luer tip and cracked spin collar.The root cause of the breakage is unknown.
|
|
Event Description
|
The affected port had been cleaned with alcohol prior to being connected.The tubing broke 16 hours later at the completion of a tpn infusion.
|
|