• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) FORTIFY ST DR, U1.6 SJ4 ID; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) FORTIFY ST DR, U1.6 SJ4 ID; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD2235-40Q
Device Problems Premature Discharge of Battery (1057); Difficult to Interrogate (1331)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/14/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that a patient presented with failure to interrogate.Different programmer did not resolve the issue.The physician alleged that the issue cased by premature battery depletion.The device is a subject to the advisory of the premature battery depletion.The device is planned to be explanted and replaced.The patient is stable.
 
Event Description
New information received indicates that the device was explanted and replaced.There was no adverse consequences to the patient before, during and after the procedure.
 
Manufacturer Narrative
Premature battery depletion was confirmed by analysis.Bench testing on the device was performed, and no sources of high current were noted.In further analysis of the battery, lithium clusters were observed, but at the time of analysis the clusters did not appear to be in a location or size that would be sufficient to cause an internal short of the battery.As a result, the cause of the premature battery depletion could not be conclusively determined.However, from these analyses, in the absence of other root causes, it is probable that the premature battery depletion was caused by a lithium cluster induced short circuit.The device is included in the premature battery depletion with implantable cardioverter defibrillator advisory notice issued by st.Jude medical on (b)(6) 2016.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FORTIFY ST DR, U1.6 SJ4 ID
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6154968
MDR Text Key61885857
Report Number2938836-2016-15934
Device Sequence Number1
Product Code LWS
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 09/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/08/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2014
Device Model NumberCD2235-40Q
Device Lot Number3939286
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/12/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/19/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/04/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age59 YR
-
-